5 Putting into operation
37
R-220 Pro Continuous Operation Manual, Version A
5 4 3
Cooling water temperature sensor
Description of function:
Enables the cooling water temperature reading at the Rotavapor
®
display
CW TEMP.
Condenser output
To sink / chiller input
Cooling water temperature sensor
Sensor socket
Installation instructions:
Use hose clamps to install the sensor at the hosing between condenser output and sink or chiller
input The sensor has no preferred flow direction
5 5
Battery driven lift function at power failure
In case of a power failure, the internal battery of the Rotavapor
®
lowers the bath This action interrupts
the heat transfer towards the product and stops the evaporation
AERATE
=
Lift
+
on
(or)
off
Store selection
X
Keep the "LIFT DOWN" button pressed while
switching on the Rotavapor® to enter the setup
menu
X
Press the "DOWN" button to toggle between 'on'
(battery drive activated) and 'off' (battery drive
deactivated)
X
Press "AERATE" button to store the selected
setting and to leave the menu