6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
87
Click on the Rack step to adapt the settings of the
rack and to add samples to the rack.
Press
SETTINGS
and adapt the settings for the
rack:
“Determine blanks first”
Yes/No
(If blanks are determined first, the risk of a cross-
contamination can be minimized for the blanks.)
“Pause after blank calculation”
Yes/No
A pause after the blank calculation leaves a user
with operator rights the chance to eliminate a faulty
blank determination from the calculation before
any sample determinations are corrected with the
calculated blank.
The third setting is only available for 2-tray samplers:
“Use tray in sampler”
A/B
Define the position of the rack in the K-377 sampler.
The step will be marked with A or B.
The positions of the rack can be filled one by one
with samples by clicking on each position.
NOTE
Using the buttons
PREVIOUS POSITION
and
NEXT POSITION
you can switch easily from one
sample to the next/previous within each parameter
screen. Thus each parameter can be adapted for all
samples of the rack in a very easy and convenient
way.