5 Putting into operation
K-375 / K-376 / K-377
Operation Manual, Version E
48
Connect the K-375 and the K-376 / K-377 with the corresponding/delivered RS232 cable (crossed)
∙ K-375: See
b
in picture 5.1
∙ K-376 / K-377: see
d
in picture 5.2
!
WARNING
Risk of burns by hot steam.
∙ Make sure to place a sample tube in the washing position(s) of the sampler.
Place an empty sample tube into the washing position of the sampler:
The washing position is on the rear right side of the tray.
For the K-376 this is the fixed position to the right of the express rack.
The K-377 provides two washing positions – one on the rear right side of each tray.
5 3 2
Connecting the transfer hoses of the K-377
K-377
∙ Connect the transfer hose and the steam
hose to the two fittings on top of the sealing
cap on the sampler arm. Secure both
connections with hose clamps
1
.
∙ The red steam hose has to be fixed to the
first position (marked with a red ring) –
pointing to the front of the instrument!
∙ Fix the plastic cable channel with the two
provided screws
2
on the sampler arm.