4 Description of function
K-375 / K-376 / K-377
Operation Manual, Version E
44
5.2 Electrical connections
5 2 1 Connections of the KjelMaster K-375
a
Power connection K-375
b
RS232 connection to K-376 /
K-377
c
RS232 connection to balance
d
LAN connection
e
USB connection to printer
f
USB connection for bar code
reader
g
Connectors for level sensors
h
Fuses (2 x 10A)
i
Connector for dosing unit (Acid)
j
Connector for additional dosing unit (Base)
k
Additional USB-ports
l
Connectors for colorimetric sensor (Ind. and Pwr. Col.) or pH
electrode (Ind. only)
m
Connectors for temperature sensor
Fig. 5.1: Electrical connections of the K-375
NOTICE
Risk of device damage by wrong voltage.
∙ Make sure that the voltage on the socket corresponds to the voltage given on
the type plate of the instrument.
∙ Always connect the device to an earthed socket. External connections and
extension cables must be provided with an earthed conductor lead (3-pole
couplings, cable or plug equipment) as the mains lead has a molded plug, thus
avoiding risks due to inadvertent defective wiring.
∙ Make sure that no electric sparks form in the device or its surroundings as they
might damage the instrument.
On the KjelMaster K-375