7 Maintenance
K-375 / K-376 / K-377
Operation Manual, Version E
140
∙ Switch both devices (the KjelMaster and the KjelSampler) off.
∙ Wait until all parts of the devices have cooled down to room temperature.
∙ Loosen the connections of the transfer hose and the steam hose on the rear side of the K-375
(see chapter 5.3.2 for details).
∙ Loosen the protective sleeve and the connections for the steam hose and the transfer hose on
the sampler arm of the KjelSampler (see chapter 5.3.1 for details).
∙ Take the transfer connection away from the devices and replace it with a new one.