6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
85
Rack 20
Enter the samples for a 20 place rack. For a 2-tray sampler the tray position (A or B)
can be selected using the
SETTINGS
button within the step.
This step can be edited within the sequence.
Pause
The sequence is stopped until it is continued by pressing
Start
.
This step cannot be changed.
Cleaning
The cleaning procedure is performed according to the settings under
System Preparation
Cleaning
Aspiration
The aspiration procedure for the sample tube and the receiving vessel is always
performed, unless the aspiration parameters of the referenced method are set to
No
.
(In this case the sample determination will stop after the determination of the sample
with the corresponding method.)
Dose H
3
BO
3
This step is thought for the protection of the electrode. 50 mL of boric acid are dosed
to the receiving vessel to keep the electrode immersed during non-usage of the
instrument.
This step cannot be changed.
Standby
The device is sent to Standby mode.
This step cannot be changed.
NOTE
The order of the steps can not be changed once they have been added to the sequence, but steps
can always be deleted and added again in a different order. The tasks Preheating, Priming, and
Cleaning are always performed “via sampler”, if used within a sequence. Even if those tasks are set
to “unit only” within the area “System Preparation” , this setting will be omitted.
Once started, all samples within a sequence are determined automatically one by one in the working-
order of the sampler. Each sequence will be deleted from the sequence list the following day, in case
all samples have been determined properly. Sequences containing faulty samples will not be deleted.
Press the
Sequences
button.
Under
Sequences
a list of all present sample
sequences for the sampler is displayed.
New sample sequences can be created with
NEW
and existing ones can be deleted, renamed or
copied.
It is also possible to import sample sequences that
have been set up on a personal computer from a
USB device or a network place.
User with administrator rights can also lock and
unlock sample sequences.
NOTE
Locked sequences can not be edited and the
determination of the sample sequence can not be
started.