BTS TMJOINT probes are internally powered.
The Probe battery replacement can be done only by BTS qualified
personnel.
The probes are sealed to avoid the access to the internal circuit
components.
Batteries are equipped with battery protection circuit to:
-over-voltage, threshold 4.3V
-under-voltage, threshold 2.8V
- short-circuits
The specific characteristic of the Wireless probes equipped with the battery
are:
Quantity:
1 per each EMG probe
Technology:
lithium polymer (Li-Poly)
Removable:
NO, BTS technical service is required