Title: New Brunswick BioFlo 3000 Bioreactor SOP

 

 

8.15.2.4.

 

Set the gases mode to DO control by selecting the box in the “Mode” row (it 
is the only box available in the “Mode” row).  Press the ALTER key until 
“DO” is displayed and then press the ENTER key. 

8.15.2.5.

 

Use the SCREEN key to choose the <Master> screen and then press the 
ENTER key. 

8.15.2.6.

 

To input the DO setpoint select the box in the “Set” row of the “DO2 1” 
column.  Use the number keypad to enter the desired setpoint, as indicated 
by the process SOP and then press the ENTER key. 

8.15.3.

 

Controlling by Agitation and % Oxygen  

Note:  In this mode agitation will be increased first until maximum rpm is reached and 
more oxygen is required, then control by % oxygen will begin. 

8.15.3.1.

 

Use the SCREEN key to choose the <Master> screen and then press the 
ENTER key. 

8.15.3.2.

 

Set the DO loop to P-I-D by selecting the box in the “Control” row of the 
“DO2 1” column. Press the ALTER key until “P-I-D” appears and then 
press the ENTER key.  

8.15.3.3.

 

Temporarily set the agitation loop to P-I-D by selecting the box in the 
“Control” row of the “Agit 1” column.  Press the ALTER key until “P-I-D” 
appears and then press the ENTER key. 

8.15.3.4.

 

To set the minimum rpm select the box in the “Set” row of the “Agit 1” 
column.  Use the number keypad to enter the desired minimum rpm, as 
indicated by the process SOP and then press the ENTER key.  

8.15.3.5.

 

Set the agitation loop to DO control mode by selecting the box in the 
“Control” row of the “Agit 1” column.  Press the ALTER key until “DO” 
appears and then press the ENTER key.  

8.15.3.6.

 

To set the maximum rpm select the box in the “Set” row of the “Agit 1” 
column.  Use the number keypad to enter the desired maximum rpm, as 
indicated by the process SOP and then press the ENTER key.   
Note: The controller will automatically adjust the rpm, therefore after the 
maximum value is entered the displayed set value may change immediately 
afterward 

8.15.3.7.

 

Press the SCREEN key and then use the arrow keys to choose the <Gases> 
screen and then press the ENTER key. 

8.15.3.8.

 

Set the gases mode to agitation and % oxygen control by selecting the box 
in the “Mode” row (it is the only box available in the “Mode” row).  Press 
the ALTER key until “Ag02DO” is displayed and then press the ENTER 
key. 

8.15.3.9.

 

Use the SCREEN key to choose the <Master> screen and then press the 
ENTER key. 

8.15.3.10.

 

Navigate to the air monitor loop by selecting the column heading “Agit 1” 
in the “Loop Name” row.  Press the ALTER key until “Air 1” appears. 

8.15.3.11.

 

Input the desired air pressure by selecting the box in the “Set” row of the 
“Air 1” column.  Use the number keypad to enter the desired air flow rate 
(usually between 1.5 and 2 L/min.) and then press the ENTER key. 

20

Summary of Contents for BIOFLO 3000

Page 1: ...llow the SOP as described and to inform the instructor about any deviations or problems that may occur while performing the procedure 4 References 4 1 New Brunswick BioFlo 3000 Bioreactor instrumentat...

Page 2: ...the DO pH and temperature cables to their appropriate outlets on the front side of the cabinet of the BioFlo 3000 8 2 2 If necessary insert the plastic plugs into the holes in the glass vessel 8 2 2 1...

Page 3: ...sion with the harvest tube 8 3 8 Secure the harvest tube in place with a nut by lowering the nut over the top of the harvest tube and hand tightening it to the threads on the topside of the headplate...

Page 4: ...the water The water valve is located under the lab bench where the BioFlo 3000 is located The pressure gauge should read between 15 and 20psi when the water is turned on 8 8 Calibrate the pH Probe 8 8...

Page 5: ...O electrolyte in the bottom housing 8 9 5 Replace the bottom housing 8 9 6 Carefully insert the DO probe into the DO port of the headplate Tighten the adaptor with a wrench 8 9 7 Ensure that the DO pr...

Page 6: ...the lab bench The pressure gauge should read between 15 and 20psi 8 12 9 Add approximately 1mL of glycerol to the thermowell and insert the RTD temperature probe into the thermowell 8 12 10 Remove the...

Page 7: ...rection Therefore the solution will flow from right to left in the tube 8 12 24 Repeat step 8 12 23 for additional feed solution flasks if needed 8 12 25 Set the pH setpoint to the desired value Refer...

Page 8: ...tlet on the front of the cabinet 8 14 1 2 Remove the protective cap from the top of the DO probe and connect the DO Cable to the DO probe 8 14 1 3 If necessary turn on the BioFlo 3000 8 14 1 4 Allow t...

Page 9: ...e Agit 1 column Press the ALTER key until P I D appears and then press the ENTER key 8 15 1 4 To set the minimum rpm select the box in the Set row of the Agit 1 column Use the number keypad to enter t...

Page 10: ...set the minimum rpm select the box in the Set row of the Agit 1 column Use the number keypad to enter the desired minimum rpm as indicated by the process SOP and then press the ENTER key 8 15 3 5 Set...

Page 11: ...then press the ENTER key 8 16 8 Use the arrow keys to select the box in the Control row of the Feed 2 column 8 16 9 Set the control mode to Level 1 by pressing the ALTER key until Lvl 1 appears and t...

Page 12: ...e 21Feb08HSA 8 17 6 7 Skip step 8 17 7 and continue with the SOP at step 8 17 8 8 17 7 To create a new recipe from the BioCommand Lite window select Create a new recipe with Batch Wizard and click Co...

Page 13: ...ck OK 8 17 7 19 8 Click 8 17 7 20 Repeat step 8 17 26 until all parameters needed are added to the Recipe B1 Trend window AGIT1 DO2 1 TEMP 1 and pH 1 should always be added 8 17 7 21 Set the graph ran...

Page 14: ...ean with a soft brush if needed Rinse thoroughly with tap water and then with deionized water 8 22 8 Remove the vessel from the heat exchanger base and pour the bleach solution down the drain Rinse th...

Page 15: ...oughly 8 22 23 Clean the heat exchanger base and rinse thoroughly 8 22 24 Ensure that all vessel parts have been cleaned and rinsed thoroughly Allow them to air dry 8 22 25 Wipe down the BioFlo cabine...

Page 16: ...nger Connectors 4 Bottom Steel Ring 5 Glass Vessel 6 Condenser Water Lines 7 Top Steel Ring 8 Agitator Shaft 9 Agitator Cover 10 Condenser Top Outlet 11 Electrical Outlet 12 Sparger Gas Port 13 Agitat...

Page 17: ...09 Page 17 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 2 Vessel Parts 1 Small Thumb Screw 2 Large Thumb Screw 3 Condenser 4 Agitator Cover 5 Agitator 6 Headplate 7 Sampler Assembly Bul...

Page 18: ...18 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 3 Headplate Underside 1 Addition Port 2 Thermowell Port 3 O ring 4 Foam Sensor Port 5 pH Probe Port 6 Agitator Shaft 7 Upper Impeller 8 L...

Page 19: ...k BioFlo 3000 Bioreactor SOP Figure 4 Assembled Headplate Top Side 1 DO Probe 2 pH Probe 3 Foam Sensor behind pH Probe 4 Condenser 5 Condenser Barbed Connectors 6 Ground Stud 7 Thermowell 8 Addition P...

Page 20: ...ve Date 20May09 Page 20 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 5 BioFlo 3000 Side View 1 Gas and Air Console 2 Cabinet 3 Power Switch 4 Feed Pump Figure 6 BioFlo 3000 Controller 1...

Reviews: