Troubleshooting
Z33146_1_001
45 / 56
9
Troubleshooting
ATM Error
• If the ATM does not work or returns an error, the ATM server software may need to be
restarted. To do this, simply quit TopSpin and restart it.
iProbe HR-MAS Errors
• If the communication with the MAS-Unit fails, reset the MAS-Unit and restart TopSpin.
• If no wobble dip is visible: Make sure the stator is not in the eject position. If it is, flip the
stator into the insert position by checking / ticking the insert button in the MAS display.
• If there is still no dip in the wobble curve, the correct routing of the current experiment
should be checked with the command
edasp
.
Problem with iProbe in TopSpin
• If a problem with TopSpin occurs during the iProbe installation, use the command
savelogs
to collect information about the installation setup and send it to your local Bruker
support.
Figure 9.1: TopSpin savelogs command
9.1
Specific for iProbes HR-MAS
Poor Line Shape
The magic angle is very critical for good shimming and line shape.
• If the line shape has deteriorated: Check the magic angle setting.
NOTICE! Sample preparation is equally critical. The presence of air bubbles inside the
rotor or leakage of the sample degrades the quality of the spectra significantly.
Summary of Contents for NMR iProbe HR-Liquids
Page 1: ...iProbes HR Liquids HR MAS User Manual Version 001 Innovation with Integrity NMR ...
Page 44: ...Maintenance 44 56 Z33146_1_001 ...
Page 46: ...Troubleshooting 46 56 Z33146_1_001 ...
Page 48: ...Replacement of Parts 48 56 Z33146_1_001 ...
Page 54: ...Contact 54 56 Z33146_1_001 ...
Page 55: ...Z33146_1_001 55 56 ...
Page 56: ...Bruker Corporation info bruker com www bruker com Order No Z33146 ...