Operation
38 / 56
Z33146_1_001
7.6.3
Variable Temperature Unit (VTU)
Since the MAS unit regulates the gas flow of the bearing gas, the VTU will only regulate the
temperature but not the flow. Therefore, the VTU must be configured to external gas flow
control in the Temperature Control Suite (VTU user interface) as highlighted in
Figure 7.19: Temperature Control Suite (VTU user interface)
7.6.4
Gradient Amplifier
In order to minimize possible artefacts coming from gradient recovery time, it is
recommended to switch on the pre-emphasis of the gradient amplifier.
Note that the pre-emphasis should be set to OFF with compensated gradient coils (typically
the case of iProbe-Liquids). Never exceed the allowed maximum current given in the TopSpin
Gradient System Parameters window.
Figure 7.20: TopSpin Gradient System Parameters.
7.6.5
Tuning & Matching
Tuning & Matching of each channel is done with the
atma
or
atmm
command (see also
). Note that in order to see the wobble curve the stator must be in the insert
position.
7.6.6
Shimming
In HR-MAS spectroscopy the rotor axis is oriented at the magic angle with respect to the
magnetic field direction. As the spinning rates are typically at least a few kilohertz, which is
much larger than the magnetic field inhomogeneity, the amplitudes of the sidebands
generated from the magnetic field inhomogeneity are small. Therefore, shimming may be
done with a restricted set of shims that is cylindrically symmetric about the MAS rotation axis
(Z', Z2', Z3', etc.). To generate these
magic angle shims
a linear combination of the standard
Summary of Contents for NMR iProbe HR-Liquids
Page 1: ...iProbes HR Liquids HR MAS User Manual Version 001 Innovation with Integrity NMR ...
Page 44: ...Maintenance 44 56 Z33146_1_001 ...
Page 46: ...Troubleshooting 46 56 Z33146_1_001 ...
Page 48: ...Replacement of Parts 48 56 Z33146_1_001 ...
Page 54: ...Contact 54 56 Z33146_1_001 ...
Page 55: ...Z33146_1_001 55 56 ...
Page 56: ...Bruker Corporation info bruker com www bruker com Order No Z33146 ...