Operation
34 / 56
Z33146_1_001
Figure 7.13: TopSpin Sample Parameters window provides the specific parameters of the sample, e.g.
liquid sample.
• The
Sample Parameters
window provides only the rotor diameter for iProbe HR-MAS.
Figure 7.14: TopSpin Sample Parameter window provides the specific parameters of the sample, e.g.
HR-MAS sample
• The peak power of the iProbe can be checked under the related submenu.
NOTICE! Never use higher power settings than the given values and use the peak
power only for short pulses, otherwise the iProbe can be damaged. Switch power
check on in order to automatically limit the peak power to these values.
Figure 7.15: TopSpin Peak Power window