Operation
Z33146_1_001
41 / 56
Figure 7.23: Standard Bore (SB) Shim System – View from Bottom
a
Projection of the
laboratory frame’s Z’-axis
φ1 Angle between X-axis Boss2/3 shim system and
projection of laboratory frame’s Z’-axis
b
X-axis Boss2/3 shim
system
φ2 Angle between X-axis Boss1 shim system and
projection of laboratory frame’s Z’-axis
c
X-axis Boss1 shim
system
7.6.6.2
Shimming in the TopShim Manual Mode
Samples with a strong signal like H
2
O are recommended for initial shimming. The sample
must be rotating at a reasonable spinning rate.
• Set all shim values to 0 (BSMS panel or using
setsh
in TopSpin).
• Load the corresponding lock solvent with
lopo
. The lock should now be in sweep mode
and the lock wiggles should be visible. Otherwise adapt Field and Gain accordingly.
However, the lock wiggles will probably be too broad for locking.
• Run
topshim changemasshims
and set the angle when asked for according to
Figure 7.24: TopShim pop-up window, where the angle phi has to be entered.
• (Use the command
topshim calreset
to reset the angle phi.)
• Optimize the Z’ shim until locking is successful by using rather large steps (e.g. 1’000
units).
Summary of Contents for NMR iProbe HR-Liquids
Page 1: ...iProbes HR Liquids HR MAS User Manual Version 001 Innovation with Integrity NMR ...
Page 44: ...Maintenance 44 56 Z33146_1_001 ...
Page 46: ...Troubleshooting 46 56 Z33146_1_001 ...
Page 48: ...Replacement of Parts 48 56 Z33146_1_001 ...
Page 54: ...Contact 54 56 Z33146_1_001 ...
Page 55: ...Z33146_1_001 55 56 ...
Page 56: ...Bruker Corporation info bruker com www bruker com Order No Z33146 ...