Installation
22 / 56
Z33146_1_001
Installation
• There are 3 pairs of equal threads
arranged opposite to each other in the
red bottom plate of the SB shim system
as marked with the arrows (1). The
quick-lock has to be mounted on one of
these pairs with the fastening screws
(M3 x 13).
• On the quick-lock a pair of opposite
arranged holes on the inner ring has to
match a pair of threads on the red
bottom plate.
• The two holes for the clip holders (1) of
the adapter determine the orientation of
the iProbe. The orientation of the iProbe
has to be set in order to attain an easy
access to the VT-gas hose. In standard
installations the front of the iProbe is
oriented towards the left when facing
the front of the magnet.
• The lever can be positioned to the right
or the left side depending on your
preference.
• Mount the quick-lock to the shim system
with the two screws M3 x 13 using the
enclosed non-magnetic Allen key.
• Make sure the Teflon ring is in place.
• The quick-lock has to be mounted level
with the shim system.