Operation
Z33146_1_001
29 / 56
7.3
ATMA
ATMA for the iProbe
The ATMA unit is integrated in the iProbe, therefore no extra unit has to be installed.
ATMA - Automatic Tuning & Matching
The iProbe can only be operated with ATMA/2, which is already integrated in AvanceNEO
spectrometers.
• After a restart of TopSpin, type the commands
edprobe
and
ii
before using atma. The
next atma command can be executed by simply typing
atma.
• Successful tuning and matching parameter will be saved automatically.
ATMM – Manual Tuning & Matching
The tuning and matching can be done by hand or via algorithms.
• To save the result of a manual tuning and matching process, an algorithm in ATMM must
be started.
• ATMM saves the channel / core combination of the last setting and therefore after a new
start the channel / core combination needs to be readjusted.