Operation
28 / 56
Z33146_1_001
7.2
Operating States of the iProbe
The meaning of the LED states and state types are listed below. The user should know them
in order to operate the system correctly and safely.
Figure 7.1: On the front of the iProbe shows an LED (1) the current status of the probe
LED
State
Meaning
Fast blinking green
Start
The probe is connected to the power line and in the
starting phase but so far not recognized by the
system.
Slow blinking green
Initialization
The probe has been recognized by the system and it
is in the initialization phase.
Slow blinking yellow
Thermal regulation
preparation
The probe will be prepared for thermal regulation.
Slow changes
yellow / green
Standby
The thermal regulation is in standby mode.
Constant yellow
Thermal regulation
The thermal regulation is running.
Blinking Red
Thermal regulation
problem
The thermal regulation failed.
Short yellow flashes
Error / Startup
The probe is connected to an auxiliary channel without
heating power. Therefore, thermal regulation cannot
run.
Table 7.1: