VPSB-DC and VPSB-DC-E
176 / 234
H172203_1_001
Ready LED
This LED is lit as soon as the FPGA program is loaded and the sensor interfaces are active
Power LED
An active Power LED indicates that the VPSB-DC(-E) is correctly powered
VT PWR 1
This LED is lit whenever the output power on the connector VTA1 is enabled
VT PWR 2
This LED is lit whenever the output power on the connector VTA2 is enabled
VTA 1 LED
Whenever a VTA is connected to the connector labeled VTA 1 and initialized correctly, the
LED will be switched on
VTA 2 LED
Whenever a VTA or BSCU is connected to the connector labeled VTA 2 and initialized
correctly, the LED will be switched on
AUX 1 LED
Whenever a VTA is connected to the connector labeled AUX 1 and initialized correctly, the
LED will be switched on
AUX 2 LED
Whenever a VTA or BSCU is connected to the connector labeled AUX 2 and initialized
correctly, the LED will be switched on
Connectors
Label
Description
Note
VTA 1
For VTA adapters that are used for power
channels (e.g. probe heater, LN2 accessory)
VTA 1 will be mapped to regulator channel 1
in Topspin
a
VTA 2
VTA 2 will be mapped to regulator channel 2
in Topspin
AUX 1
For VTA adapters that do not require heater
power (e.g. BCU control)
AUX 1 will be mapped to auxiliary channel 1
in Topspin
b
AUX 2
AUX 2 will be mapped to auxiliary channel 2
in Topspin
a
If a 2nd VPSB-DC(-E) is in use then its VTA 1 will be mapped to regulator channel 3 and VTA 2 to
regulator channel 4 respectively.
b
If a 2nd VPSB-DC(-E) is in use then the device at AUX 1 will be mapped to auxiliary channel 3 and
AUX 2 to auxiliary channel 4 respectively.
Table 15.8: VPSB-DC(-E) Front Panel Connectors
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...