VTA
182 / 234
H172203_1_001
Bruker
Part
Number
Name
Marking
Typical usage
[Power or
Auxiliary]
a
Typical usage
Connected devices
Z119237
BSMS/2 VTA TC-2T
TC-2T
POWER
RT, Solids and Flow
Probes
VTN/WVT/DVT
Z116924
BSMS/2 VTA BTO
BTO
POWER
RT probes with BTO2000
Z116923
BSMS/2 VTA CRP
CRP
POWER
CryoProbe
Z120851
BSMS/2 VTA FLOW-NMR
FLOW-NMR
POWER
Flow NMR Capillary
Heater
Z120728
BSMS/2 VTA TC-2E
TC-2E
POWER
Probes for solids, high
temperature
Z119238
BSMS/2 VTA LN2
LN2
POWER
N2 Evaporator
N2 heat exchanger
Z116925
BSMS/2 VTA BCU
BCU
AUX
BCU-05, BCU-X
a
POWER means that this application needs heater power and therefore must be connected through
a cable that contains heater power wires to an output of the VARIABLE POWER SUPPLY BOARD
(VPSB-DC(-E)).
Table 16.1: List of Available VTAs
16.3
Technical Data
VTA TC-T
Parameter
Min
Type
Max
Unit
Temperature measurement
Range
-270
400
°C
Thermocouple type T
Accuracy
+/- 0.5
°C
Number of channels
1
Connector type
a
A
Heater
Connector type
H
Maximum voltage
50
V
Maximum current
7
A
Safety temperature measurement
Range
-200
+850
°C
Thermocouple type K
Resolution
0.1
Measurement update rate
1
s
-1
a
Table 16.2: VTA TC-T
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...