SPB
H172203_1_001
159 / 234
14.5.7
Auxiliary Bus Connector (VT Accessory, SPB-E)
With introduction of the Bruker Sample & Variable Temperature System (BSVT) a new
generation of sensor interface adaptors are available. These adapters convert sensor signals
into a digital data-stream. These sensors are typically connected to the AV4 VARIABLE
POWER SUPPLY BOARD DC (VPSB-DC).
The SPB-E variant is equipped with one supplementary auxiliary bus interface connector.
The interface is hot-plug capable, has automatic connect/disconnect detection and power
supply signals are short-circuit proof.
14.5.8
Pneumatics
For detailed technical specifications please see
.
NOTICE
Never supply the console with non-filtered gas.
Gas supply filter 88437 must be installed!
The SPB contains all valves, valve drivers, and pneumatic connectors necessary to control:
• Sample lift
• Sample spinner (sample rotation)
• VT gas
• Probe flush gas
• Shim system cooling gas
• Emergency lift detection
The SPB replaces the former BSMS/2 PNK3, PNK3S and PNK5 boards.
The valve drivers are galvanically isolated and the module is controlled by the ELCB which
also saves the calibration values.
To provide usage of different gases for sample transport/shim cooling and delicate probe
head gases (VT, spin, flush gas), these two functional groups have separate gas inputs.
For reasons of sample safety it is necessary to connect a buffer to the lift system.
With the exception of the lift system and emergency lift gas for the CryoProbe sample safety
option, there are no further calibration procedures necessary. In particular spin calibration is
not necessary.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...