SCB20
82 / 234
H172203_1_001
• Are codes 255 only followed by further codes 255?
• Are all codes except 255 different from each other?
• Apart from values of 255, are the values of the codes decreasing from left to right?
2. Uncoded BOSS1 Shim Systems with less than 20 Shims
The ELCB firmware checks if the number of connected loads corresponds with the number of
shim coils defined in the BOSS file. For BOSS1 systems, the ELCB assumes 20 loads.
However, there are older, uncoded Shim Systems with less than 20 loads (e. g. only 17 shim
coils). After power up, the user is informed about that by an error message. On the BSMS
Service Web there is a button for confirmation of the actual number of loads of the currently
connected Shim System. Afterwards, the load check is executed accordingly.
3. Check ftp.bruker.ch for New BOSS files
New BOSS files are published on the following ftp location:
ftp://ftp.bruker.ch/NMR/download/servtools/bsmstool/boss/
Check this directory (and the sub-directories) for new BOSS files and installation guides/
service information. It might be necessary to update your locally installed BOSS files.
4. Swap Cables if You Have a
Plug
>Type Shim System
The two Shim cables 'A' and 'B' could be connected in wrong order. Swapping the connectors
at one side would eliminate the error. Make sure that the BSMS is switched off before
swapping the cables!
In case of a defective Shim cable, swapping of the cables at both ends would move a wrong
code to another position, e. g. from {<correct> <correct> <not correct>} to {<not correct>
<correct> <correct>}.
Plug or unplug this connector only when the system is powered off.
5. Perform a Detailed Hardware Check
Ask your local Bruker office for assistance.
Summary of Contents for NMR AV4 BSMS System
Page 1: ...BSMS System for AVANCE NEO User Manual Version 001 Innovation with Integrity NMR...
Page 10: ...Contents x H172203_1_001...
Page 22: ...Safety 22 234 H172203_1_001...
Page 26: ...Transport Packaging and Storage 26 234 H172203_1_001...
Page 42: ...Chassis Mainframe 42 234 H172203_1_001...
Page 46: ...Fan Tray 46 234 H172203_1_001...
Page 55: ...ELCB H172203_1_001 55 234 Figure 8 1 ELCB Front Panel with LED s and Connectors...
Page 69: ...ELCB H172203_1_001 69 234 Figure 8 13 Lock RF Boards Diagnostics...
Page 70: ...ELCB 70 234 H172203_1_001...
Page 120: ...L TRX L 19F 120 234 H172203_1_001...
Page 148: ...BSVT Concept 148 234 H172203_1_001...
Page 166: ...SPB 166 234 H172203_1_001...
Page 172: ...VPSB DC and VPSB DC E 172 234 H172203_1_001 Figure 15 3 Block Diagram of the VPSB DC E...
Page 180: ...VPSB DC and VPSB DC E 180 234 H172203_1_001...
Page 187: ...VTA H172203_1_001 187 234 Figure 16 1 VTA Cable Connectors...
Page 193: ...VTA H172203_1_001 193 234 16 7 Ordering Information See Basic BSVT Configuration 124...
Page 194: ...VTA 194 234 H172203_1_001...
Page 200: ...Nitrogen Level Sensor 200 234 H172203_1_001...
Page 204: ...Radiation Shield Temperature Monitoring MAG RS 204 234 H172203_1_001...
Page 208: ...Installation and Initial Commissioning 208 234 H172203_1_001...
Page 210: ...Operation 210 234 H172203_1_001...
Page 216: ...Replacement of Parts 216 234 H172203_1_001...
Page 222: ...Contact 222 234 H172203_1_001...
Page 226: ...List of Figures 226 234 H172203_1_001...
Page 229: ...Glossary H172203_1_001 229 234 Glossary...
Page 230: ...Glossary 230 234 H172203_1_001...
Page 232: ...Index 232 234 H172203_1_001...
Page 233: ...H172203_1_001 233 234...
Page 234: ...Bruker Corporation info bruker com www bruker com Order No H172203...