Running the minispec mq20 Toothpaste Analyzer
E1400912_1_001
39
After the first sample is done, go on with further calibration samples. By default the
software requests a minimum of 3 samples per calibration parameter. However often the
calibration can be improved by including further samples, for example 5 or 6.
When the measurement of all the samples is completed, switch to the calibration results.
Afterwards move forward to the calibration line and inspect the correlation and other
parameters describing the calibration quality.