Technical Data
E1400912_1_001
21
4.5
Specifications mq20 Toothpaste Analyzer
4.5.1
Weights and Dimensions
Electronic Control Unit
Magnet Unit
Dimensions (W x H x D) in cm
23 x 31 x 60
47 x 31 x 60
Weight (kg)
20
85
Table 4.5: Weights and Dimensions for mq20 Toothpaste Analyzer
4.5.2
Electrical Specifications
minispec mq-Series systems (wideline versions / see corresponding label on back of the unit):
The instruments (wideline) are specified for the following mains voltages:
Input Voltage Range: 100 – 240 V AC 50/60 Hz
Fuse rating mq-Series NF: 2 x 3.15 AT
All Systems
Grounding is vital for the safe operation of the instrument. Never operate with the grounding
conductor disconnected. If the local wiring does not provide a ground, an earth-to-chassis
connection must be added at the user’s location.
Note
: Since the input voltage may be unstable in a laboratory environment, it is
recommended to connect the minispec to a separate circuit that is not used by any other
equipment. Furthermore, a voltage stabilizer or UPS may be required to avoid instrument
instabilities or other unexpected system behavior caused by a fluctuating input voltage.
4.6
Rating Plate
The rating plate is located at the power input and includes the following information:
Figure 4.1: The minispec MQ-Series Rating Plate
Summary of Contents for minispec mq20
Page 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Page 5: ...Contents E1400912_1_001 5 Index 83 ...
Page 6: ...Contents 6 E1400912_1_001 ...
Page 14: ...Safety 14 E1400912_1_001 ...
Page 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Page 22: ...Technical Data 22 E1400912_1_001 ...
Page 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Page 76: ...Maintenance 76 E1400912_1_001 ...
Page 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Page 80: ...Contact 80 E1400912_1_001 ...
Page 82: ...List of Figures 82 E1400912_1_001 ...
Page 85: ...E1400912_1_001 85 ...
Page 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...