Running the minispec mq20 Toothpaste Analyzer
E1400912_1_001
35
During Daily Check, the system runs a number of tests, all following one after the other.
If all the tests passed, the software will indicate this accordingly. The corresponding message
will also be displayed in the main menu.
7.3
Calibrations
TD-NMR is a secondary technique and as such the Toothpaste Analyzer needs to be
calibrated with reference standards.
For all other applications the customer will need to calibrate the system with their own
reference standards. Typically 4-6 standards with known reference values are needed per
calibration line. In most cases the mq20 Toothpaste Analyzers are calibrated in a classical
way, i.e. the NMR signals of FID or Hahn-echo (or both) are considered and the
corresponding amplitudes are correlated with the sample properties.
For different types of samples (like different types toothpaste or mouthwash) the user may
need to create different calibration lines, one line per type of sample. It is recommended to
name the calibrations accordingly.
Summary of Contents for minispec mq20
Page 1: ...minispec mq20 Toothpaste Analyzer User Manual Version 001 Innovation with Integrity AIC ...
Page 5: ...Contents E1400912_1_001 5 Index 83 ...
Page 6: ...Contents 6 E1400912_1_001 ...
Page 14: ...Safety 14 E1400912_1_001 ...
Page 18: ...Transport Packaging and Storage 18 E1400912_1_001 ...
Page 22: ...Technical Data 22 E1400912_1_001 ...
Page 72: ...minispec exe Software for Diagnostic and Service Purposes 72 E1400912_1_001 ...
Page 76: ...Maintenance 76 E1400912_1_001 ...
Page 78: ...Dismantling and Disposal 78 E1400912_1_001 ...
Page 80: ...Contact 80 E1400912_1_001 ...
Page 82: ...List of Figures 82 E1400912_1_001 ...
Page 85: ...E1400912_1_001 85 ...
Page 86: ...Bruker Corporation info bruker com www bruker com Order No E1400912 ...