70
ZTKS0172 / Z31964 / 04
Technical Data MS 500‘89 Ascend
9.9
Fringe Field Plot
Figure 9.8:
Fringe field plot of the magnet system at maximum field
Fringe Field
Unit
R max
Unit
Z max
Unit
200
mT
0.2
9
m
0.49
m
5.0
mT
0.50
m
0.88
m
3.0
mT
0.54
m
0.97
m
1.0
mT
0.62
m
1.21
m
0.5 (5 Gauss)
mT
0.70
m
1.40
m
0.2
mT
0.92
m
1.73
m
0.1
mT
1.17
m
2.07
m
0.05
mT
1.52
m
2.49
m
max. magnetic field B0 at access point
314
mT
max. field gradient dB/dz at access point
4.
7
T/m
Table 9.12: Fringe field data of the magnet system at maximum field
This p
age
wi
ll b
e
rep
lace
d
by spe
cific
te
chn
ical
d
at
a.
Summary of Contents for 500'89 Ascend Aeon
Page 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Page 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Page 8: ...8 ZTKS0172 Z31964 04 Contact ...
Page 28: ...28 ZTKS0172 Z31964 04 Safety ...
Page 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Page 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Page 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Page 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Page 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Page 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Page 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Page 98: ...98 ZTKS0172 Z31964 04 Index ...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...