46
ZTKS0172 / Z31964 / 04
Troubleshooting
6.2.4
During Energizing and Shimming
Indicator
Possible reason
Solution
Performed by
The current lead
can not be inserted
completely into the
connector.
The connector is
covered with ice
(frozen moisture or
nitrogen ice).
Carefully remove the ice
with warm helium gas.
Use the dipstick or the
precooling tube as tubing for
the warm helium gas to
remove small ice spots.
Bruker Service
The shorting plug
was not removed.
Remove the shorting plug
with the shorting plug tool.
Bruker Service
The orientation of
the current lead is
not correct.
Carefully reposition the
current lead until it can be
inserted correctly into the
connector.
Bruker Service
Main coil heater
test fails.
Power supply
defective.
Replace the power supply.
Bruker Service
Connector or
cables defective.
Clean connectors or replace
cables, if necessary.
Bruker Service
Setting of sense
voltage fails.
The main coil
heater switch is
"OFF".
The main coil
switch is not
opened.
Switch the main coil heater
to "ON" and check the main
coil heater current to be
adjusted correctly.
Bruker Service
The main coil
heater current is
not correct. The
main coil switch is
not opened.
Adjust main coil heater
current correctly.
Bruker Service
The auxiliary
shorting plug is
inserted in the
current lead turret
by mistake and
makes a short
circuit across the
main coil.
Remove the auxiliary
shorting plug and insert it in
the helium fill-in turret.
Bruker Service
Summary of Contents for 500'89 Ascend Aeon
Page 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Page 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Page 8: ...8 ZTKS0172 Z31964 04 Contact ...
Page 28: ...28 ZTKS0172 Z31964 04 Safety ...
Page 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Page 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Page 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Page 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Page 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Page 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Page 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Page 98: ...98 ZTKS0172 Z31964 04 Index ...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...