64
ZTKS0172 / Z31964 / 04
Technical Data MS 500‘89 Ascend
9.4
Safety Valves
Figure 9.3:
Safety valves of the cryostat
1 One-way valve ot the nitrogen
vessel
2 Safety valve of the nitrogen
vessel
3 One-way valve of the helium
vessel
4 Quench
valve
Opening pressure
mbar
psi
One-way valve of the nitrogen vessel
10
0.15
Safety valve of the nitrogen vessel
140
2
One-way valve of the helium vessel
15
0.2
Quench valve
180
2.6
Table 9.6: Opening pressure of the safety valves of the magnet system
Summary of Contents for 500'89 Ascend Aeon
Page 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Page 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Page 8: ...8 ZTKS0172 Z31964 04 Contact ...
Page 28: ...28 ZTKS0172 Z31964 04 Safety ...
Page 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Page 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Page 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Page 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Page 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Page 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Page 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Page 98: ...98 ZTKS0172 Z31964 04 Index ...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...