48
ZTKS0172 / Z31964 / 04
Troubleshooting
Table continued
Indicator
Possible reason
Solution
Performed by
Shim current can
not be set correctly.
The control cable is
not connected
correctly to the
current lead or to
the power supply.
Connect the control cable
correctly to the current lead
and to the power supply.
Bruker Service
The switch „Main
Coil/OFF/Shim
Coil“ is in the wrong
position.
Change the switch position.
Bruker Service
Shims do not affect
the NMR signal.
Shim heater
current is not
correct. The shim
switches are not
opened.
Set the shim heater current
to the specified value
(See
).
Bruker Service
Magnet system
does not reach
specification.
Magnetic material
inside RT bore
tube.
Carefully clean the RT bore
tube.
Bruker Service
Large ferromag-
netic parts near the
magnet system.
1.
As much as possible,
keep all ferromagnetic
parts away from the
magnet system.
2.
Repeat shimming.
Bruker Service
Summary of Contents for 500'89 Ascend Aeon
Page 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Page 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Page 8: ...8 ZTKS0172 Z31964 04 Contact ...
Page 28: ...28 ZTKS0172 Z31964 04 Safety ...
Page 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Page 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Page 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Page 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Page 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Page 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Page 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Page 98: ...98 ZTKS0172 Z31964 04 Index ...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...