Technical Data MS 500‘89 Ascend
9.10
Filling Volume, Evaporation Rate and Hold Time
Cryogenic Consumption during Installation
The consumption of liquid cryogenic agents during installation consists of consumption
for cooling down the cryostat, for energizing, cryo shimming and quench reserve.
1
Maximum time interval between two fillings
Cryogenic Agents Consumption
Value
Unit
Nitrogen
600
l
Helium (needed for cool down)
300
l
Helium (needed for energizing, cryo shimming and quench
reserve)
300
l
Table 9.13: Cryogenic agents consumption during installation
Cryogenic Agents
Value
Unit
Nitrogen vessel total volume
134
l
Nitrogen refill volume
104
l
Nitrogen evaporation rate
270
ml/h
Nitrogen hold time
1
16
days
Helium vessel total volume
90
l
Helium refill volume
68
l
Helium evaporation rate
19
ml/h
Helium hold time
1
150
days
Helium refill volume after quench
(cool down and refill)
150
l
Table 9.14: Cryogenic agents
This p
age
wi
ll b
e
rep
lace
d
by spe
cific
te
chn
ical
d
ata
.
Summary of Contents for 500'89 Ascend Aeon
Page 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Page 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Page 8: ...8 ZTKS0172 Z31964 04 Contact ...
Page 28: ...28 ZTKS0172 Z31964 04 Safety ...
Page 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Page 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Page 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Page 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Page 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Page 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Page 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Page 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Page 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Page 98: ...98 ZTKS0172 Z31964 04 Index ...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...