Chapter 5 – Maintenance and Troubleshooting
Warnings and Error messages
58
Brüel & Kjær
Type 2239 B
Hand-Arm Vibration Functions User Manual
5.4
Warnings and Error messages
If you get an error message while using or calibrating your
Type 2239 B, then refer to the chart below.
Message
Occurs
Solution
If you press
,
, , ,
or
while the instru-
ment is in the middle
of a measurement.
Wait for measure-
ment to finish or
press
to
stop
measurement and
then make your
settings.
During calibration, if
foreign vibrations
have corrupted the
calibration signal or if
the calibrator is not
switched on.
Make sure you
have turned on and
connected the cali-
brator correctly
a
and/or move to a
more vibration-free
location. If the
error persists, then
contact your
Brüel& Kjær dealer.
a. See section 3.1 for more information.
During calibration, if
the instrument has
discovered that the
required calibration
factor is more than
±
1.5 dB from the fac-
tory calibration.
Check the condi-
tions listed above.
If the error per-
sists, then you may
need a new fac-
tory calibration.
Contact your
Brüel& Kjær dealer
for assistance.
*** Warning ****
Function Not
Available While
Measuring
***** Error ****
Calibration
Stopped due to
Input Error
***** Error ****
New Initial
Calibration
Needed