Chapter 6 – Specifications
Specifications
62
Brüel & Kjær
Type 2239 B
Hand-Arm Vibration Functions User Manual
Fig.6.1 Frequency response with linear frequency weighting
10
100
1k
Frequency (Hz)
10k
100k
1
0.01
10
1
10
Acceleration m/s
2
960562e
6.1
Specifications
STANDARDS:
Conforms with ISO 8041 Type 2 and ISO
5349
INPUT:
0.36 pC/ms
-2
for accelerometer Type 4505
FREQUENCY WEIGHTINGS:
Linear (Unweighted): see Fig.6.1
Hand-Arm Vibration Weighting: see
Fig.6.2
MEASURING RANGES:
Hand-Arm: 8 – 1000 Hz.
Linear: 8 – 5000 Hz.
Inst, Low Range Setting: 0.1 – 316 m/s
2
Inst, High Range Setting: 1 – 3160 m/s
2
Peak, Low Range Setting: 0.14 – 447.2 m/
s
2
Peak, High Range Setting: 1.4 – 4472 m/s
2
DETECTORS:
RMS Averaging Time: 1 s
Peak Rise Time: <100
µ
s
Automatic reset at 1 s intervals.
PARAMETERS:
A
min
, A
max
, A
eq
, A
eq4
, and A
eq8
are cal-
culated based on 1 s exponential averag-
ing of the instantaneous RMS readings
(Inst). A
mp
is the highest peak reading
(Peak).
OVERLOAD INDICATION
Instantaneous indication of overload and
latched overload. Stored records also in-
clude a latched overload indicator.
DISPLAY:
4 line LCD showing:
• Input signal level – indicated with a
quasi-analogue bar (updated 15 times
per second)
• Selected parameters with level
• Warnings for: battery low and overload
• Measuring range
• Elapsed measurement time
• Menus for displaying and editing set-
tings
• Recalled records
Features a back-light for easy viewing,
which can be turned on and off and in-
cludes an auto time-out to save batteries.
bb122411.book Page 62 Tuesday, December 11, 2001 2:34 PM