Brooks Automation
7. Troubleshooting
Part Number: 319430 Rev. A
Error Messages
Error
Code
Cause
Corrective Action
E163
Caps on Pins detected, RECOVERY
MODE activated.
See
Use "EJECT CAPS" operation to push caps off
machine.
Use "INITIALIZE" to restart the instrument.
E164
Timeout stage sequence in decap
operation.
Call service
E165
Timeout opening safety door during
manual retry decap operation.
Check the door has not been bent out of shape
and remove any object that is blocking operation.
Call service.
E166
Timeout homing stage sequence in
manual retry decap operation.
Call service.
E167
Timeout positioning stage sequence in
manual retry decap operation.
Call service.
E238
Emergency stop is active.
Twist release emergency stop.
Power cycle unit.
Call service.
Copyright © 2019, Brooks Automation, Inc.