Brooks Automation
7. Troubleshooting
Part Number: 319430 Rev. A
Error Recovery
Error Recovery
Error
Symptom
Resolution
CAP
ERROR
Tube is not de-capped properly, the IntelliXcap will
automatically make a second attempt.
If the IntelliXcap fails on the second attempt, an error
message is shown on the screen, and the IntelliXcap
stops.
Manually add a new cap to the
tubes and perform a new de-
capping cycle.
RECAP
ERROR
(Error
Code:
136)
Cap is improperly placed onto the corresponding
tubes during the recapping process.
Select the
Initialization
(Restart)
button and start the
IntelliXcap.
Table 7-1: Typical Errors
Manual Recovery
In any error situation, you have the option to cancel the process. You will be prompted to start a
manual recovery process. Choose the most relevant case available on the screen.
Use the up and down arrows to access additional options.
Step
Action
1.
Use the Access door Up-option to lift the access door;
2.
Try to move the Screwing Head Up;
3.
If there are still caps attached to the ejecting pins, use a bowl to collect the falling caps
and then choose Eject Caps from the menu;
4.
Once the caps have been ejected and collected, use the option Open Tray.
Copyright © 2019, Brooks Automation, Inc.