7. Troubleshooting
Part Number: 319430 Rev. A
Error Messages
Error
Code
Cause
Corrective Action
E137
Cartridge has reached lower operational
limit.
Power cycle instrument.
Call service.
E138
Timeout opening stage/nest in Open Tray
operation.
Call service.
E139
Cartridge eject initialized, no cartridge
detected at startup.
Run cartridge load command.
Call service.
E140
Safety door has been forced out of open
position.
Verify there is no objects blocking the safety
door.
Run a decap / recap cycle.
Call service.
E141
Safety door has been forced out of close
position.
Verify there is no objects blocking the safety
door.
Power cycle the unit.
Run a decap / recap cycle.
Call service.
E142
Object detected in cartridge eject
sequence.
Remove any objects on the stage.
Call service.
E143
Object not detected in cartridge load
sequence.
Ensure cartridge is placed on the stage.
Call service.
E144
Cartridge height detect wrong during
cartridge load
Ensure you have the right cartridge for the right
instrument.
Call service.
E145
Lightcurtain calibration fault in decap
sequence.
Remove any direct light sources.
Call service.
E146
Lightcurtain calibration fault in recap
sequence.
Remove any direct light sources.
Call service.
E147
Lightcurtain calibration fault in cartridge
eject sequence.
Remove any direct light sources.
Call service.
E148
Lightcurtain calibration fault in cartridge
load sequence.
Remove any direct light sources.
Call service.