3. Specifications and Site
Requirements
IntelliXcap
Site Requirements
Part Number: 319430 Rev. A
Electrical Requirements
The system must only operate with the power supply and frequency specified on the system
identification stickers mounted on the side of the device. Operating the system with any other
power supply or frequency can result in damage to the equipment.
Parameter
Specification
Supply Voltage
100-120 VAC 1/N/PE / 220-240 VAC 1/N/PE
Use IEC 320 plugs only
Ground must be connected at all times
Maximum Power Consumption
500W
Idle Power Consumption
100W
Supply Frequency
The machine operates below the noise emissions level: < 70 dB(A)
Fuses
Two fuses:
100-120V AV 5A fuse
220- 240 V AC 5A fuse
Insulation Resistance
Not less than 1MΩ at 1,000V
Phase 1 = 50GΩ
Neutral = 50GΩ
UI Connection
RS 232 cable
Table 3-4: Electrical Requirements
15
Copyright © 2019, Brooks Automation, Inc.