Trouble
Possible Cause
Check/Corrective Action
Output stays at 0 Volts
Clogged Sensor.
Clean sensor. Refer to cleaning procedure. Section 4-2.
regardless of flow.
Output signal stays at +6.8V
Defective PC Board.
Replace PC Board. Refer to Section 4-4.
and there is no flow through
the meter.
Defective Sensor.
Replace Sensor Assembly. Refer to Section 4-4.
Meter grossly out of calibration. Partially clogged sensor.
Clean sensor, refer to the cleaning procedure. Section 4-2.
Flow is higher than indicated.
Meter grossly out of calibration. Partially clogged restrictor.
Replace restrictor. Refer to Section 4-4.
Flow is lower than indicated.
Meter output oscillates.
Anticipate potentiometer out of adjustment. Adjust anticipate potentiometer. Refer to Section 3-4.
Faulty pressure regulator.
Check regulator output.
Defective PC Board.
Replace PC Board. Refer to Section 4-4.
Table 4-1 Bench Troubleshooting
Table 4-2 Sensor Troubleshooting