Section 1 Introduction
Electrical Specifications
FLOW INPUT (VOLTAGE) SPECIFICATIONS
FLOW INPUT (CURRENT) SPECIFICATIONS
FLOW OUTPUT (VOLTAGE) SPECIFICATIONS
FLOW OUTPUT (CURRENT) SPECIFICATIONS
ANALOG I/O ALARM OUTPUT*
ANALOG I/O VALVE OVERRIDE SIGNAL SPECIFICATIONS**
Communication Protocol
RS485
Profibus
®
DeviceNet
TM
Electrical Connection
1 x 15-pin Male Sub-D, 1 x 15-pin Male Sub-D/ 1 x M12 with
(A)
1 x 9-pin Female threaded coupling nut
Sub-D
(B)
Analog I/O
0-5 V, 1-5 V, 0-10 V, N/A
0-20 mA, 4-20 mA
Power Max./Purge
From +13.5 Vdc to From +11 Vdc to
+27 Vdc
+25 Vdc
Power Requirements Watts, Max.
Valve Orifice > 0.032”: 8 W Valve Orifice > 0.032”: 10 W
Valve Orifice
≤
0.032”: 5 W Valve Orifice
≤
0.032”: 7 W
Without Valve: 2 W Without Valve: 4 W
Nominal Range
0-5 Vdc, 1-5 Vdc or 0-10 Vdc
Full Range
(-0.5) -11 Vdc
Absolute Max.
18 V (without damage)
Input Impedence
>990 kOhms
Required Max. Sink Current
0.002 mA
Nominal Range
4-20 mA or 0-20 mA
Full Range
0-22 mA
Absolute Max
.
24 mA (without damage)
Input Impedence
100 Ohms
Nominal Range
0-5 Vdc, 1-5 Vdc or 0-10 Vdc
Full Range
(-1)-11 Vdc
Min Load Resistance
2 kOhms
Nominal Range
0-20 mA or 4-20 mA
Full Range
0-22 mA (@ 0-20 mA); 3.8-22 mA (@ 4-20 mA)
Max. Load
380 Ohms (for supply voltage: < 16 Vdc)
Type
Open Collector
Max. Closed (On) Current
25 mA
Max. Open (Off) Leakage
1
µ
A
Max. Open (Off) Voltage
30 Vdc
Floating/Unconnected
Instrument controls valve to command set point
VOR < 0.3 Vdc
Valve Closed
1 Vdc < VOR < 4 Vdc
Valve Normal
VOR > 4.8 Vdc
Valve Open
Input Impedence
800 kOhms
Absolute Max. Input
(-25 Vdc) < VOR < 25 Vdc (without damage)
*The Alarm Output is an open collector or "contact type" that is CLOSED (on) whenever an alarm is active.
The Alarm Output may be set to indicate any one of various alarm conditions.
** The Valve Override Signal (VOR) is implemented as an analog input which measures the voltage at the input and controls the
valve based upon the measured reading as shown in this section.
EtherNet/IP
TM
1x 5-pin M8 Male Nano
Change Connector /
2x 4-pin M12 Female D
Coded Connector
N/A
From +13.5 Vdc to
+27 Vdc
Valve Orifice > 0.032”: 11 W
Valve Orifice
≤
0.032”: 7 W
Without Valve:3W
Web-based Network Settings Interface
N/A
N/A
Network Configuration
is DHCP. Network address
is 192.168.1.100
Table 1-2 SLAMf Series Electrical Specifications
3
1
Sanitary fittings - Model code 5A, 5B, 5C, 5D & 5E rated to 500 psi Maximum Pressure
²
600 lpm of H2 possible with decreased accuracy. Greater than 40 psig inlet required for flows greater than 100 lpm N
₂
equivalent
³
1000 psi/70 bar for UL Certificate
⁴
4500 psi/310 bar available as a special on SLAMf61 only