Installation and Operation Manual
X-VA-GT1000-eng
Part Number: 541B078AAG
April, 2018
GT1000 Series
Section 2 Installation
17
3. Slide the metering tube up behind the plastic tube retainer until it clears
the lower end fitting. Hold the inlet (lower) float stop in position when
removing the tube. When the tube is clear of the lower end fitting pull out
on it until you can remove it from the upper end fitting.
4. Remove the four socket head screws on the rear of the body casting,
then remove the end fittings and their clamps.
5. Drill four 5/16 holes in the body casting at the cast-in drill marks.
6. Reinstall the end fittings and their clamps. Align horizontal end fittings
with the piping before tightening the clamp bolts to 95 inch-pounds (10.7
Newton Meter).
7. Bolt the meter to the panel using four 1/4-20 or M6 bolts of a proper
length for the panel thickness.
8. Slide the metering tube onto the upper end fitting. Again, be careful not to
allow the float to fall out of the metering tube. Insure that the plastic tube
retainer is pulled all the way out. Refer to Figure 1-3 to insure that the
orientation of the float is correct and that the inlet and outlet float stops
are in position.
9. Push the metering tube in and down to seat it on the lower end fitting.
10.Push the plastic tube retainer in to hold the tube in position.
11.Install the front shield on the meter and secure with four screws tightened
to 35 inch-pounds (4 Newton Meters).