42 Cleaning the iSleep 20 and Replacement of Accessories
iSleep 20 clinician’s manual
Doc. 003923 En-Us V-2
Always clean the parts or use a new set when used by a new patient.
All parts that come into contact with the respiration gas must be cleaned as fol-
lows:
1
Place the dismantled parts in hot water containing washing-up liquid.
2
Remove fouling with a brush.
3
Rinse parts thoroughly under running hot water.
4
Shake water out of all parts.
5
Dry the parts completely.
6
Store in dust-free location.
Check the patient circuit for damage regularly. In case of damage, replace the
circuit.
If the parts need to be disinfected, this can be done in a bath of Virkon
®
or Lyse-
tol
®
Med, for instance. Then rinse the parts well in clear water and dry them
thoroughly.
9.2 Cleaning and Replacing the Patient Air Filter
Washable Filter (grey color, mandatory)
Replace the washable filter at least once a year. Wash the filter at least once a
week.
1
Wash the filter using warm water and a mild detergent.
2
Rinse thoroughly.
3
Dry the filter by squeezing it out in a towel. Do not wring the filter.
Disposable Filter (white color, optional)
The patient air filter is located on the
side of the iSleep 20. There are two
types of filters: