8
9
English
The production date is given by the LOT number and can be deciphered as explained below:
The fi rst 3 numeric digits in the LOT number represent the day of the year of manufacture.
The next 2 numeric digits represent the last two numbers of the calendar year of manufacture and
the letter(s) at the end designate the manufacturer of the product. (E.g.: LOT No.: 12313tav this
product was made on the day 123, year 2013 at the manufacturer code tav.)
Errors and troubleshooting
Product specifi cations
Error message
Situation
Solution
When 20% of battery life is left, the
display fl ashes the low battery warning
symbol; however, the device can still
work until the battery life has 0% left.
Replace batteries.
If the steady battery icon is
the only symbol shown on the display,
the device cannot work. The battery
should be replaced immediately.
Replace batteries.
This message displays when the
measured temperature is lower than
34.4 °C (93.9 °F) or higher than 42.2 °C
(108 °F) or when the room temperature is
outside the operating range of
15 °C – 40 °C (59 °F – 104 °F).
Re-measure the temperature, carefully
following the instructions in “How to use”
section.
Blank display. Thermometer does not
have power.
Please check if the batteries have been
loaded correctly. Also check polarity (<+>
and <–>) of batteries.
Contact customer service if thermometer
still does not function.
Type:
No touch + forehead thermometer (NTF 3000)
Measuring range:
34.4 °C – 42.2 °C (93.9 °F – 108 °F)
Resolution:
0.1 °C (0.1 °F)
Laboratory Accuracy:
±0.2 °C for the range 35 °C – 42 °C (±0.4 °F for 95 °F – 107.6 °F)
±0.3 °C (outside this temperature range)
(Ambient Temperature: 15 °C to 40 °C (59 °F to 104 °F))
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
Normal temp range = single long beep for 2 sec duration
Fever = Red or Yellow temp range: 10 short beeps
Operating temperature:
15 °C – 40 °C (59 °F to 104 °F)
Automatic Switch-off :
Approx. 60 seconds after last measurement has been taken
Weight:
99.5 g (with batteries), 77.1 g (w/o batteries)
Long term storage ranges
Storage/transport temperature:
-25 °C to 60 °C (-13 °F – 140 °F)
Humidity:
15–95% non-condensing
Battery:
(2) AA Batteries - at least 1000 measurements
Pressure:
700-1060 hPA (0,7-1,06 atm)
This product conforms with the provisions of the European directive 93/42/EEC (Medical Device
Directive). This product carries the CE mark and is manufactured in conformity with the RoHS
Directive 2011/65/EU.
This device conforms to the following standards:
EN 60601-1 Medical electrical equipment. General requirements for basic safety and
essential performance
EN 80601-2-56 Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
EN 12470-5 «Clinical thermometers» – Part 5: Performance of infrared ear thermometers (with
maximum device)
EN 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral Standard: Electromagnetic compatibility – Requirement and
tests
NOTE: Do not use this device in the presence of electromagnetic or other interference outside the
normal range specifi ed in EN 60601-1-2.
Guarantee
EMC Tables and Consumer Card are available on our website at www.hot-europe.com/support.
For support requests, please visit www.hot-europe.com/support or fi nd service contact
information on the last page of this manual.
UK only: This guarantee in no way aff ects your rights under statutory law.
Calibration
This product has been calibrated at the time of manufacture. If used according to the use
instructions, periodic calibration is not required. If at any time, you question the measuring
accuracy, please contact our service representative (see last page for contact information) or visit
www.hot-europe.com/support.
This device is not intended as a substitute for regular check-ups by your doctor, please consult your
doctor if you have any doubt about the temperature reading.
LOT and Serial numbers of your product must be provided as they are essential to record and follow
up your inquiry or claim.
The LOT and Serial numbers are located in the battery compartment.