94
Q32: If I have some configuration problems, which files I have that to send to the supplier?
A32: Both Recorder.cfg. and IO.dat. files in C:/Observer/Historical are required under this case.
Q33: Does the recorder have an option to supply DC power supply to the transmitters?
A33: The 24VDC auxiliary power supply card AP181 can offer the power to 6 pieces of transmitters.
Q34: I am trying to connect recorder on RS 232/RS422/RS485/Ethernet but I couldn’t set up the
communication properly.
A34:
Please refer detailed information to
5.1 Observer I & Observer II PC software guide, 5.2
Ethernet Configuration
&
5.3 RS 232, RS485, RS422 Configuration.
Q35: When export data to Excel CSV files, what if the historical data is not appearing in spread
sheet?
A35: Make sure that the selected pens are available at destination list. If they are available at source list,
then move them to destination list, and then you can export data. Also make sure that data is
available at the selected period or interval time. Otherwise, you may not see anything in the target
Excel sheet.
Q36: I want to count pulses to measure power consumption in KWH. Each pulse is one KWH.
Can I use Math application for this?
A36: Yes, use DI and configure Events of Channel as follows
Event Type = High, Job 1 = Increase Counter, Target = Counter 1
But please note that maximum value counter can store is 65535.
Please note that recorder can detect maximum of 30 pulses/Minute at a frequency of 0.5 Hz. The
better possible solution is to use an external converter for the application. The standard pulses to
4-20mA converters are available in the market where you can set pulse width and duration
between pulses.
Q37: I am getting the following message in Real time viewer. What to do next?
A37: The firmware V2.21 or earlier version does not allow the user to key in special characters like “
.
”
“
,
” “
*
” “
+
” “
-
” etc. in Tag name. The user may change the tag name, or update the recorder
firmware and PC software to V2.31 to avoid error message above.
Summary of Contents for VR06
Page 1: ...User Manual Paperless Recorder VR06 UMVR061D Brainchild Electronic Co Ltd 4th Edition 10 2013 ...
Page 17: ...17 Figure 2 6 Figure 2 7 Figure 2 8 ...
Page 19: ...19 Figure 2 10 Figure 2 11 ...
Page 25: ...25 Digital input card DI181 Figure 2 21 Analog output card AO183I AO183V Figure 2 22 ...
Page 27: ...27 Figure 2 25 Figure 2 26 ...
Page 69: ...69 Make sure that RS 232 communication setting at the recorder and PC are equal ...
Page 86: ...86 7 Trouble shooting ...
Page 87: ...87 ...
Page 88: ...88 ...