90
Q6: What kind of security is available in recorder?
A6: There are two kinds of security. The normal security offers a common password. The high security
with FDA 21 CFR part 11 compliance is available from version 2.2 on both recorder and PC
software Observer. It supports maximum 30 users with different passwords. The administrator has
the right to define supervisor and operators. The supervisor can change configuration, save and clear
data but not operators. Operators can only view the historical data, events and status.
Q7: What is Boot Rom? Anything related to Math option? Is it possible to upgrade it into new
version via internet or CF card?
A7: Boot ROM is a chip available in recorder, which contains boot loader information. The Boot ROM
BT181 has no option. The Boot ROM BT182 has options of Math, Counter, Totalizer & FDA 21
CFR part 11 compliance. It is required to remove BT181 away and install BT182 on the socket
when those options are needed. It is not possible to upgrade it into new version via internet or CF
card.
Q8: How does it upgrade into the new version of recorder system and PC software?
A8: If necessary, contact your supplier about the latest updates. The supplier is able to supply the latest
version of recorder system and PC software via email or FTP ( file transfer protocol ) for easy
download through internet. Follow the procedure to update it into the new version of recorder
system. Install the new version of Observer software. It shall automatically uninstall the old one.
Q9: If it needs to upgrade the recorder system, what do we have to take care?
A9: Any recorder system before V 1.20, it is mandatory to update first to V 1.20 and then to the latest
version i.e. it needs to upgrade twice. Direct update to the latest version could damage the DOC.
Q10: What are the differences between Observer I & Observer II PC software?
A10: Observer I is supplied for free. Observer II is supplied at additional cost. Observer I is used for
recorder configuration from PC and also it is used for reading historical data in PC at set time
period of 1 to 24 hours. Observer II is used for real time data monitoring in PC along with
configuration and historical data view.
Q11: How many AI, AO, DI, DO cards, and 24VDC auxiliary power card can be inserted into the
recorder?
A11: Maximum 6 cards in any combination can be inserted into rear Slots.
Q12: What are those larger capacities of compact flash cards supported by the recorder and
recommended manufacturer?
A12: We already offer the maximum 2GB CF card with each recorder. To be fully compatible, we
usually recommend
SanDisk
brand.
Summary of Contents for VR06
Page 1: ...User Manual Paperless Recorder VR06 UMVR061D Brainchild Electronic Co Ltd 4th Edition 10 2013 ...
Page 17: ...17 Figure 2 6 Figure 2 7 Figure 2 8 ...
Page 19: ...19 Figure 2 10 Figure 2 11 ...
Page 25: ...25 Digital input card DI181 Figure 2 21 Analog output card AO183I AO183V Figure 2 22 ...
Page 27: ...27 Figure 2 25 Figure 2 26 ...
Page 69: ...69 Make sure that RS 232 communication setting at the recorder and PC are equal ...
Page 86: ...86 7 Trouble shooting ...
Page 87: ...87 ...
Page 88: ...88 ...