2.5.1. Surgical environment:
prevention of explosions and ignition
Argon itself is not flammable, but the high temperature of the plasma can ignite
other flammable materials and gases.
For this reason, always purge endangered intestinal sections with argon or
CO
2
gas before activating the HF current.
Do not use the argon device in areas where there is a risk of explosion.
Do not use any flammable or explosive liquids.
Stop using the argon device if any display components fail.
Avoid using ignitable anaesthetics and gases which support combustion (e.g.
nitrous oxide or oxygen) during surgery in regions such as the head or thorax.
Use only non-flammable cleaning agents, disinfectants and solvents (for
adhesives). If you use flammable cleaning agents, disinfectants or solvents,
ensure that they have fully evaporated before using the HF surgical
equipment.
Ensure that no flammable liquids collect beneath the patient or in body
cavities (e.g. the vagina). Suction and/or purge body cavities before activating
the device.
Wipe off all liquids before using the HF device.
Ensure that no endogenous gases are present that could ignite.
Ensure that all materials saturated with oxygen (e.g. cotton or gauze) are kept
far enough away from the HF environment that they cannot ignite.
Always check oxygen lines for leaks.
If the output power is not sufficient with the customary settings, first ensure that:
the neutral electrode is correctly attached;
the working electrodes are clean;
the connectors are properly connected;
before selecting a higher output power.
Do not use the device if even the smallest irregularity is present.
2.5.2. Application of the neutral electrode
Observe the instructions for use of the neutral electrode in the operating
instructions of the HF device and the instructions on the packaging of the neutral
electrode.