4 Preparation
26
ARC PLUS Operating Manual
900-001_IFU_V1.0_11270-S0-20130118-EN
Figure 4-4:
Before purging
6. Acknowledge the display by touching the Purge button.
– or –
All illuminated indicators except the Purge button go dark after 10 seconds.
Purge the system with argon gas by touching the Purge button.
The argon device is ready for use.
Figure 4-5:
Purging active and Purge button actuated
7. Before starting surgery, check the fill level of the gas bottle on the pressure
gauge of the pressure reducer, or on the status indicator on the HF generator
as described in the operating instructions for the HF generator (e.g. "Argon
menu" section for the ARC 400).
The electronic pressure sensor reports the fill level, which is shown on the
display of the associated HF generator (e.g. ARC 400). If the fill level of the
bottle is adequate, the surgery can begin.
8. A new gas bottle must be connected when a message appears on the display
of the HF device indicating that the argon pressure is less than 30 bar. If a
second gas bottle is connected, changeover to the second gas bottle takes
place automatically. The purpose of this monitoring function is to ensure that
an adequate gas supply is available.
The argon device is de-activated automatically when an argon mode is not
selected on the HF generator. The operating mode indicator
2
changes from green
(Standby) to orange (Sleep).
NOTE
Observe the manufacturer’s warnings and instructions for the accessories
before connecting them to the device.