7 Cleaning
36
ARC PLUS Operating Manual
900-001_IFU_V1.0_11270-S0-20130118-EN
7.
Cleaning
7.1.
Cleaning of accessories
Clean the accessories (e.g. surgical handpieces, instruments, active
electrodes, neutral electrodes and cables) as described in the associated
operating manuals.
Check the accessories before and after use for damage and ensure that they
are working properly.
7.2.
Disinfection and cleaning
NOTE
Incorrect use of the HF device can cause damage to the unit
Never sterilize the argon device. Instead, clean or disinfect it.
WARNING
Risk of electric shock and fire
Unplug the power cable before cleaning the device.
For cleaning surfaces, use the approved cleaning agents or disinfectants only
as specified by the manufacturer.
Follow the procedure specified in your hospital or use a proven infection
control method.
Ensure that no liquid penetrates the device.
1. Apply the cleaning and/or disinfection agent (with the device switched off and
the power cord unplugged).
2. Wipe the agent off with a sponge or cloth moistened with clean water.
3. Dry the device using a clean, lint-free cloth.
Do not sterilise the pressure reducer or the gas hose. Use wiping disinfection to
clean and disinfect the surfaces.