![Boston Scientific Latitude 3300 Operator'S Manual Download Page 72](http://html.mh-extra.com/html/boston-scientific/latitude-3300/latitude-3300_operators-manual_2796085072.webp)
66
Table 5.
Guidance and Manufacturer’s Declaration—Electromagnetic
Compatibility
(continued)
The LATITUDE Programming System, Model 3300, is suitable for use in
professional health care facility environments. The customer or the user of this
system should assure that it is used in such an environment.
Test
Compliance
Electromagnetic
environment—guidance
Voltage dips
0%
U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0%
U
T
for 1 cycle and 70%
U
T
for 25/30 cycles at 0°
Voltage interruptions
0%
U
T
for 250/300 cycles
a.
Voltage dips and interruptions:
U
T
is the AC mains voltage prior to application of the test level.
LATITUDE PROGRAMMING SYSTEM SECURITY
Sensible security practices are needed to protect patient data and the
LATITUDE Programming System integrity when connected to a network. The
Programmer incorporates features that facilitate management of network
security. These features work in conjunction with the security practices of
hospitals and clinics to provide safe and secure operation of the Programmer
and protect the attached network.
NOTE:
All patient data is encrypted on the Programmer internal hard drive,
and the Programmer has network security safeguards in place to prevent
malicious attacks.
Software
All installed software has been approved by Boston Scientific and general
purpose software installation is not permitted. This minimizes the potential for
vulnerabilities to be exposed. Internal software that runs the Programmer is
locked from change and is re-verified upon each execution. Whenever Boston
Scientific software updates are available, install them as soon as possible.
Programmer settings should only be modified per guidance from verified
Boston Scientific technical support or Health Care Delivery personnel.
Patient Data Management
Refer to the
Patient Data Management Operator’s Manual (Model 3931)
for
additional security information.
Network
Refer to the
Network and Connectivity Operator’s Manual (Model 3924)
for
additional networking and connectivity security information.