9
•
Inductive telemetry.
Using the Programmer on battery power only may
reduce the telemetry distance (from wand to implanted device). If needed,
use AC power to improve inductive telemetry.
•
Battery operation during long-term storage .
Remove battery to prevent
discharging when storing the Programmer for long periods (e.g., months).
•
Date and time accuracy.
Inability to access a remote time server could
lead to discrepancies in the Programmer time. As a backup, the Boston
Scientific representative can set the time and date manually.
•
Patient data.
Patient data may be stored on the Programmer up to 14
days and appropriate precautions should be taken to secure the
programmer from unauthorized access.
•
Delete all patient data from the Programmer (refer to the
Patient Data
Management Operator’s Manual (Model 3931)
for delete instructions)
before shipping the Programmer or at any time when the Programmer
leaves your direct control.
•
Only connect to known Bluetooth
®
devices to reduce the potential of
transmitting patient data to inappropriate printers or devices.
•
USB devices.
USB devices connected to the Programmer should be
controlled to limit the potential introduction of malware.
•
External device battery usage.
Using external devices (USB, display
monitor) will deplete the battery. To extend Programmer performance,
refrain from using external devices when on battery power only and the
battery level indicator shows 25% or less remaining.
•
Software.
Ensure that you have the latest software versions installed (see
"Software Update Tab" on page 35). As a backup, your local Boston
Scientific representative can provide software updates using a USB pen
drive.
•
Model 6395 Telemetry Wand shipped non-sterile.
The Model 6395
Telemetry Wand is shipped non-sterile. Remove the wand from all
packaging material before sterilizing it. If the wand is to be used in a sterile
field, it must be actively sterilized before use or enclosed in a disposable
sterile surgical sheath (Model 3320) during use. Refer to "Cleaning the
Programmer and Accessories" on page 43 for sterilization and cleaning
information.
•
Model 3203 S-ICD Telemetry Wand shipped non-sterile .
The Model
3203 S-ICD Telemetry Wand is shipped non-sterile. Remove the wand
from all packaging material before use. If the wand is to be used in a sterile
field, it must be enclosed in a sterile intraoperative probe cover (Model
3320) during use. Refer to "Cleaning the Programmer and Accessories" on
page 43 for cleaning information.
•
Electrical and magnetic interference.
Avoid establishing telemetry
communication between the Programmer and the PG when the
Programmer is in close proximity to monitors, high-frequency
electrocautery equipment, or strong magnetic fields. The telemetry link
may be impaired.