64
Services and must accept interference that may be caused by such stations,
including interference that may cause undesired operation. This transmitter
shall be used only in accordance with the FCC Rules governing the Medical
Device Radiocommunication Service. Analog and digital voice communications
are prohibited. Although this transmitter has been approved by the Federal
Communications Commission, there is no guarantee that it will not receive
interference or that any particular transmission from this transmitter will be free
from interference.
CAUTION:
No modification of this equipment is allowed unless approved by
Boston Scientific. Changes or modifications not expressly approved by Boston
Scientific could void the user’s authority to operate the equipment.
The electromagnetic emissions and immunity information is provided in Table 5
Guidance and Manufacturer’s Declaration—Electromagnetic Compatibility on
page 64.
Table 5.
Guidance and Manufacturer’s Declaration—Electromagnetic
Compatibility
The LATITUDE Programming System, Model 3300, is suitable for use in
professional health care facility environments. The customer or the user of this
system should assure that it is used in such an environment.
Test
Compliance
Electromagnetic
environment—guidance
Protection of radio
services and other
equipment
CISPR 11
Group 1
Class A
The LATITUDE
Programming System,
Model 3300, uses RF
energy only for its
intended uses in
communication with the
implanted device or
connectivity functions. Its
RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
Protection of the public
mains network
CISPR 11 Class A
IEC 61000-3-2
IEC 61000-3-3
The LATITUDE
Programming System,
Model 3300, is suitable for
use in professional health
care facility environments.
Electrostatic Discharge
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, and
± 15 kV air
Radiated RF EM field
3 V/m from 80 MHz to 2.7
GHz
Proximity fields from RF
wireless communications
equipment
380 - 390 MHz: 27 V/m
430 - 470 MHz: 28 V/m
704 - 787 MHz: 9 V/m
800 - 960 MHz: 28 V/m