54
Table 3.
Possible Causes and Corrective Actions for LATITUDE Programming
System Problems
(continued)
Symptom
Possible Cause
Corrective Action
Use another Model 3300
Programmer or Model
6395 Telemetry Wand.
Repeat interrogation.
If this does not correct the
issue, contact Boston
Scientific using the
information on the back
cover of this manual.
Telemetry RF signal
obstructed
Assure that a clear line-of-
sight path exists between
the LATITUDE
Programming System and
PG. Repeat interrogation.
Telemetry RF signal
interference
Reposition LATITUDE
Programming System.
Unplug USB devices.
Repeat interrogation.
RF Telemetry fails
Reposition the Model
6395 Telemetry Wand
over the transvenous PG
and repeat interrogation.
Use the Model 3203 S-
ICD Telemetry wand as an
additional antenna.
LATITUDE Programming
System software version
not current
Contact Boston Scientific
using the information on
the back cover of this
manual.
Noise problems: ECG
Improper patient
connections
Recheck patient leads for
adequate skin contact and
correct limb lead
placement. Confirm Right
Leg Drive is connected.
Consult ECG textbooks for
additional ECG
techniques.
Excessive radio emissions
from equipment
Check surrounding area
for electrical equipment
that is powered on and not
needed. Move unneeded
equipment away from
patient and/or LATITUDE
Programming System, or