![Boston Scientific Latitude 3300 Operator'S Manual Download Page 34](http://html.mh-extra.com/html/boston-scientific/latitude-3300/latitude-3300_operators-manual_2796085034.webp)
28
•
PSA STAT PACE - when a PSA session has been enabled, it configures
the PSA with STAT PACE settings and functionality.
NOTE:
All emergency function commands prompt the user to exit and start a
new session.
Figure 11.
STAT Button Pop-up in a High-voltage PG Session with the PSA
Application Running
The top row buttons (PG STAT PACE, DIVERT THERAPY, and STAT SHOCK)
display during a PG session.
Figure 12.
PSA STAT PACE Button Pop-up Outside Any PG session with the PSA
Application Running
If not in a PG session, the STAT button brings up the following dialog with no
buttons: "There is no active device session. Press “Interrogate” to initiate the
Quick Start
™
function."
If in a PSA session only (no PG interrogated), then the same dialogue displays
along with the PSA STAT PACE button (see Figure 12 PSA STAT PACE Button
Pop-up Outside Any PG session with the PSA Application Running on page
28).