![BOREAL LASER GasFinder2 Operation Manual Download Page 61](http://html1.mh-extra.com/html/boreal-laser/gasfinder2/gasfinder2_operation-manual_2788300061.webp)
9.11
Boreal Laser Inc.
GasFinder2 Operation Manual — Appendices
July 17th 2017
Appendix I — Log to Memory
Log to Memory Menus:
To use this option, enter the
Menu
by pressing
Ent
on the keypad.
The display will read
Concentration Units
. Press the up arrow
fi ve times and the
Log to
Memory Menus
will be displayed.
Press
Ent
and
Log Buffer Mode
will be displayed.
Buffer Mode
:
Pressing
Ent
again will enable the user to choose between
fi ll
and
wrap
mode. The data
can be stored in the memory in a
wrap
- around mode where, when the memory is full, data
overwrites the initial records. Thus only the last readings(131,072) are retained. Alternatively,
if the
fi ll
mode is used, once 131,072 readings have fi lled the memory, all future data are
discarded. Pressing
ESC
will return to
Log Buffer Mode
, and then pressing the down arrow
will lead to the next item.
Log to Memory
:
When this is
Enabled
, the internal logging option is activated.
Log
Memory Records
: This item lists the number of records currently in the memory. The storage
capacity is 131,072 lines. Recording a data point every second will give over 36 hours of
recording. If the average is set to one minute using the
Average Window
set to 58 sec, then
the recording time is nearly 3 months. Note that a dual gas system with have half the storage
time available because of twice as many data points.
Log Memory Dump
:
Pressing
Ent
will download the stored data to a computer terminal through the serial data
port. The data can be captured using
GasViewOP
or
Hyperterminal
.
See the description of
HyperTerminal
on page 4.4.
Downloading the data will not erase the data in the internal memory. It can be downloaded
as many times as required, provided it has not been cleared.
Clear Memory Log
:
Pressing
Ent
will erase all data stored in the memory. Switching the
GasFinder2
off does not
erase the data. Data can only be erased using
Clear Memory Log
.
To return to the normal operating mode, press
ESC
twice.