BIRTHNING BED AVE 2 SERVICE MANUAL
Date
06-2017
Version
1.2
6.3.1.
Disassembling the Lift Column
Before disassembling the lift column, you must first remove the supporting area; see
1. Disassemble the left LED lighting (Picture 6.3.1.1.)
2. Disconnect all the connectors going to the control box, with the exclusion of the
power-supplying cable and the Nurse Call cable going to the connector from behind
on the frame of the bed. After that, pull all the cables through the opening in the frame
(Picture 6.3.1.2)
Picture 6.3.1.1
Picture 6.3.1.2
3. Unbolt the interface cover from beneath of the upper frame (Picture 6.3.1.3)
4. Unbolt the cable-attaching part (Picture 6.3.1.4) and disconnect all the connectors of
the interface.
Picture 6.3.1.3
Picture 6.3.1.4